| Model |
PC-818 A/So |
PC-816 A/So |
PC-708 |
PC-815 |
PC-812 |
PC-350 |
PC-320 |
| Sample |
PC 818-05
0.5ml X 54 tubes
PC818-02
0.2 ml X 96 tubes
|
PC 816-05
0.5ml X 54 tubes
PC 816-02
0.2ml X 96 tubes
|
PC 708-05
0.5ml X 30 tubes
PC 708-02
0.2ml X48 tubes |
0.5ml X
54 tubes |
0.2ml X
96 tubes |
0.5ml X
18tubes |
0.2ml X
32 tubes |
| Lid system |
Auto-pressure lid / Adjustable
pressure lid |
Top-down compression heated lid |
| Temperature control |
Microprocessor based PID control |
| Heating and Cooling |
Controlled by solid state module
(Peltier)
|
Block temperature
range |
0.0oC to
99.9oC |
3.0oC to
99.9oC
(Room temperature
25oC) |
| Block accuracy |
±0.1oC |
Block uniformity |
Less than ±0. 5oC between holes
|
Gradient set temperature |
5oC to 20oC |
________________ |
| Temperature ramping |
Heating
rate:<3oC/sec,average faster than
1o C/sec - Cooling rate:<2 o C/sec
average faster than 1o C/sec |
Control mode |
Programmed Temp Mode |
Constant Temp mode/Program Temp mode |
| Setting times |
1sec to 99min 59 sec, HOLD,α |
| Maximum number of cycle |
99 cycles / 1 segment |
| Reaction temperature step |
Max 25 step (5 step / seg X 5 segments) |
| Program files |
Max 160 files
(60+100 on memory card) |
Max 45 files (15 files x 3 boxes) |
Cover heater control |
Block interlock, Fixed temperature, Heater off |
Block interlock |
| Display |
| 6-inch high resolution colored LCD |
Wide-format LCD |
W134.0 x H40.4LCD, (8 x 40 letters) |
|
|
| Interface |
| Printer, RS-485, Analog 5ch |
|
|
|
External Dimensions
WxDxH |
W282mm x
402mm x
H280mm |
W282mm x
D402mm
x
H280mm |
W282mm x D402mm x H280mm |
W234mm x
D370mm
x H58mm |
| Power supply |
AC200V to
240V,
50-60Hz,
400W |
AC200V to
240V,
50-60Hz,
350W |
AC200V to
240V,
50-60Hz,
600W |
AC200V to
240V,
50-60Hz,
350W |
AC200Vto
240V,
50-60Hz,
250W |
| Weight |
10kg |
8.5kg |
10kg |
5.5kg |
*A&S denotes Automatic
and Adjustable Models respectively |
*The 384 well plate model is factory option (part no PC 818 -384 AL/PC 816-384AL |
*Specificatons may be changed without notice |
Constant Temp Mode :1) LIGATION -16oC, 2) ENZYME REACTION-37oC, 3) DENATURATION- 95oC, 4) COOLING-4oC 5) USER SETTING
Maintain Samples at 4oC on completion of reaction. |
|
|
| |